Using double diffusion encoding it is possible to acquire microstructural and water exchange information. Here, simulations are used to study how restriction effects influence apparent exchange measurements. The simulations indicate that at the chosen experimental parameters the restriction effect can be considerable for large pores and small mixing times, τm. In typical exchange rate experiments using clinical MR systems with τm > 40 ms, the restriction effect can probably be neglected if pores are small.
Originalsprache | Englisch |
---|---|
Seiten | 1085 |
Seitenumfang | 1 |
Publikationsstatus | Veröffentlicht - 27.04.2017 |
Veranstaltung | ISMRM 25th Annual Meeting & Exhibition - Honolulu, USA / Vereinigte Staaten Dauer: 22.04.2017 → 27.04.2017 |
Tagung, Konferenz, Kongress
Tagung, Konferenz, Kongress | ISMRM 25th Annual Meeting & Exhibition |
---|---|
Land/Gebiet | USA / Vereinigte Staaten |
Ort | Honolulu |
Zeitraum | 22.04.17 → 27.04.17 |