Analysis of PMMA versus CaP titanium-enhanced implants for cranioplasty after decompressive craniectomy: a retrospective observational cohort study

Dominik Wesp, Harald Krenzlin, Dragan Jankovic, Malte Ottenhausen, Max Jägersberg, Florian Ringel, Naureen Keric*

*Korrespondierende/r Autor/-in für diese Arbeit

Abstract

Numerous materials of implants used for cranioplasty after decompressive craniectomy (DC) have been investigated to meet certain demanded key features, such as stability, applicability, and biocompatibility. We aimed to evaluate the feasibility and safety of biocompatible calcium-phosphate (CaP) implants for cranioplasty compared to polymethylmethacrylate (PMMA) implants. In this retrospective observational cohort study, the medical records of all patients who underwent cranioplasty between January 1st, 2015, and January 1st, 2022, were reviewed. Demographic, clinical, and diagnostic data were collected. Eighty-two consecutive patients with a mean age of 52 years (range 22–72 years) who received either a PMMA (43/82; 52.4%) or CaP (39/82; 47.6%) cranial implant after DC were included in the study. Indications for DC were equally distributed in both groups. Time from DC to cranioplasty was 143.8 ± 17.5 days (PMMA) versus 98.5 ± 10.4 days (CaP). The mean follow-up period was 34.9 ± 27.1 months. Postoperative complications occurred in 13 patients with PMMA and 6 in those with CaP implants (13/43 [30.2%] vs. 6/39 [15.4%]; p = 0.115). Revision surgery with implant removal was necessary for 9 PMMA patients and in 1 with a CaP implant (9/43 [20.9%] vs. 1/39 [2.6%]; p = 0.0336); 6 PMMA implants were removed due to surgical site infection (SSI) (PMMA 6/43 [14%] vs. CaP 0/39 [0%]; p = 0.012). In this study, a biocompatible CaP implant seems to be superior to a PMMA implant in terms of SSI and postoperative complications. The absence of SSI supports the idea of the biocompatible implant material with its ability for osseointegration.

OriginalspracheEnglisch
ZeitschriftNeurosurgical Review
Jahrgang45
Ausgabenummer6
Seiten (von - bis)3647-3655
Seitenumfang9
ISSN0344-5607
DOIs
PublikationsstatusVeröffentlicht - 12.2022

Fördermittel

TrägerTrägernummer
Johannes Gutenberg-Universität Mainz

    UN SDGs

    Dieser Output leistet einen Beitrag zu folgendem(n) Ziel(en) für nachhaltige Entwicklung

    1. SDG 3 – Gesundheit und Wohlergehen
      SDG 3 – Gesundheit und Wohlergehen

    DFG-Fachsystematik

    • 2.23-07 Klinische Neurologie, Neurochirurgie und Neuroradiologie

    Fingerprint

    Untersuchen Sie die Forschungsthemen von „Analysis of PMMA versus CaP titanium-enhanced implants for cranioplasty after decompressive craniectomy: a retrospective observational cohort study“. Zusammen bilden sie einen einzigartigen Fingerprint.

    Zitieren