A high power driving and selection field coil for an open MPI scanner

G. Bringout, H. Wojtczyk, W. Tenner, M. Graeser, M. Grüttner, J. Hägele, R. Duschka, N. Panagiotopoulos, F.M. Vogt, J. Barkhausen, T.M. Buzug

Abstract

Three Litz wires have been used in parallel in order to decrease the layer to layer voltage down to 4 kV and the coil voltage down to 8 kV. A two millimeter thick epoxy plate has been glued between each layer, in order to prevent any electrical breakdown between the layers. A special epoxy has been used to minimize the inner temperature of the coil, which reached 180°C with a dissipated power of 5.2 kW. The use of parallel Litz wire leads to currents having different phase angle in every wire which has to be corrected.
OriginalspracheEnglisch
Titel 2013 International Workshop on Magnetic Particle Imaging (IWMPI)
Seitenumfang1
Herausgeber (Verlag)IEEE
Erscheinungsdatum01.03.2013
Seiten1-1
ISBN (Print)978-1-4673-5520-9
ISBN (elektronisch)978-1-4673-5522-3
DOIs
PublikationsstatusVeröffentlicht - 01.03.2013
Veranstaltung3th International Workshop on Magnetic Particle Imaging (IWMPI 2013) - Berkeley, USA / Vereinigte Staaten
Dauer: 23.03.201324.03.2013

Fingerprint

Untersuchen Sie die Forschungsthemen von „A high power driving and selection field coil for an open MPI scanner“. Zusammen bilden sie einen einzigartigen Fingerprint.

Zitieren